/usr/share/doc/libsane/umax/sane-umax-config-doc.html is in libsane-common 1.0.22-7ubuntu1.
This file is owned by root:root, with mode 0o644.
The actual contents of the file can be viewed below.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141 142 143 144 145 146 147 148 149 150 151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169 170 171 172 173 174 175 176 177 178 179 180 181 182 183 184 185 186 187 188 189 190 191 192 193 194 195 196 197 198 199 200 201 202 203 204 205 206 207 208 209 210 211 212 213 214 215 216 217 218 219 220 221 222 223 224 225 226 227 228 229 230 231 232 233 234 235 236 237 238 239 240 241 242 243 244 245 246 247 248 249 250 251 252 253 254 255 256 257 258 259 260 261 262 263 264 265 266 267 268 269 270 271 272 273 274 275 276 277 278 279 280 281 282 283 284 285 286 287 288 289 290 291 292 293 294 295 296 297 298 299 300 301 302 303 304 305 306 307 308 309 310 311 312 313 314 315 316 317 318 319 320 321 322 323 324 325 326 327 328 329 330 331 332 333 334 335 336 337 338 339 340 341 342 343 344 345 346 347 348 349 350 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365 366 367 368 369 370 371 372 373 374 375 376 377 378 379 380 381 382 383 384 385 386 387 388 389 390 391 392 393 394 395 396 397 398 399 400 401 402 403 404 405 406 407 408 409 410 411 412 413 414 415 416 417 418 419 420 421 422 423 424 425 426 427 428 429 430 431 | <!DOCTYPE HTML PUBLIC "-//W3C//DTD HTML 4.01 Transitional//EN">
<html>
<head>
<meta http-equiv="Content-Type" content="text/html; charset=iso-8859-1">
<meta name="Author" content="Oliver Rauch">
<meta name="Description" content="unix scanner driver for UMAX scanners">
<title>SANE-umax - Configuration</title>
</head>
<body text="#000000" bgcolor="#FFFFFF" link="#0000EF" vlink="#51188E" alink="#FF0000">
<table>
<tr>
<td width=200>
<img SRC="sane-logo.jpg" alt="SANE-logo" hspace=60 align=BOTTOM>
<br>
<img SRC="sane-umax-text.jpg" alt="UMAX-logo" HSPACE=50>
</td>
<td>
<blockquote>
<h1><font color="#009900">Configuration</font></h1>
</blockquote>
</td>
</tr>
</table>
<hr WIDTH="100%">
<table>
<tr>
<td VALIGN=TOP WIDTH="250" BGCOLOR="#E0E0FF">
<br>
<ul>
<li><a href="sane-umax-doc.html">Index</a></li>
<li>
Frontend options:
<ul>
<li><a href="sane-umax-standard-options-doc.html">Standard options</a></li>
<li><a href="sane-umax-advanced-options-doc.html">Advanced options</a></li>
</ul>
</li>
<li>
Supported scanners:
<ul>
<li><a href="sane-umax-powerlook-doc.html">Power Look series</a></li>
<li><a href="sane-umax-mirage-doc.html">Mirage series</a></li>
<li><a href="sane-umax-astra-doc.html">Astra series</a></li>
<li><a href="sane-umax-vista-doc.html">Vista series</a></li>
<li><a href="sane-umax-uc-doc.html">UC series</a></li>
<li><a href="sane-umax-others-doc.html">others</a></li>
<li><a href="sane-umax-not-listed-doc.html">Not listed scanners</a></li>
<li><a href="sane-umax-scanner-clones-doc.html">UMAX scanner clones</a></li>
</ul>
</li>
<li><a href="sane-umax-parport-doc.html">UMAX parallel port scanners</a></li>
<li><a href="sane-umax-config-doc.html">Configuration</a></li>
<li><a href="sane-umax-speed-doc.html">Speed of UMAX scanners</a></li>
<li><a href="http://www.rauch-domain.de/sane-umax">SANE-umax homepage</a></li>
<li><a href="http://www.sane-project.org">SANE homepage</a></li>
<li><a href="http://www.xsane.org/">XSane homepage</a></li>
</ul>
<br>
</td>
<td VALIGN=TOP BGCOLOR="#FFFFFF">
<blockquote>
<br>
<h2>
CONFIGURATION:
</h2>
<blockquote>The configuration file for this backend resides in
<dl>
<dd>
/usr/local/etc/sane.d/umax.conf.
</dd>
</dl>
</blockquote>
<blockquote>
Its contents is a list of options and device names that correspond to UMAX
scanners. Empty lines and lines starting with a hash mark (#) are ignored.
A sample configuration file is shown below:
<br>
<dl>
<dd>option scsi-buffer-size-min 65536</dd>
<dd>option scsi-buffer-size-max 131072</dd>
<dd>option scan-lines 40</dd>
<dd>option preview-lines 10</dd>
<dd>option scsi-maxqueue 2</dd>
<dd>option handle-bad-sense-error 0</dd>
<dd>option execute-request-sense 0</dd>
<dd>option force-preview-bit-rgb 0</dd>
<dd>option slow-speed -1</dd>
<dd>option care-about-smearing -1</dd>
<dd>option calibration-full-ccd -1</dd>
<dd>option calibration-width-offset -1</dd>
<dd>option calibration-bytes-pixel -1</dd>
<dd>option invert-shading-data -1</dd>
<dd>option lamp-control-available 0</dd>
<dd>#scsi Vendor Model Type Bus Channel ID LUN</dd>
<dd>scsi UMAX * Scanner * * * * *</dd>
<dd>/dev/scanner</dd>
<dd># this is a comment</dd>
<dd>/dev/sge</dd>
</dl>
</blockquote>
<blockquote>
The special device name must be a generic SCSI device or a
symlink to such a device. To find out to which device your scanner is assigned
and how you have to set the permissions of that device, have a look at
<a href="http://www.sane-project.org/man/sane-scsi.5.html">sane-scsi</a>.
</blockquote>
<blockquote>
When the generic scsi driver of the system supports selection
of scsi buffer sizes the umax backend calculates the scsi buffer
sizes in dependance of the numbers of bytes per scanline.
With <i>scsi-buffer-size-min</i> and <i>scsi-buffer-size-max</i>
the scsi buffer size can be defined in a range between 4KB and 1MB.
Especially the minimum value is very important. If this value is
set too small the backend is not able to send gamma tables to the
scanner or to do a correct color calibration. This may result in
strange color effects. If the minimum value is set too large then
the backend is not able to allocate the requested scsi buffer size
and aborts with out of memory error. The default is 32KB, for
some scanners it should be increased to 64KB.
<p>
The option <i>scan-lines</i> defines the the number of lines that
shall be scanned with one scsi read command for a real scan.
The number of lines that are read with one scsi command for
preview scans are defined by the option <i>preview-lines</i>.
Both have to be in the range between 1 and 65536.
<p>
With the option <i>scsi-maxqueue</i> you can define the number
of scsi commands that shall be queued. UMAX scanners do not
support command queueing - as far as I know. But some scsi controller
drivers do support queueing in the kernel. In this case it does not
make sense to queue more than 2 commands.
<p>
When you are able to do color previews but color scans do not
work you can set the option <i>force-preview-bit-rgb 1</i>.
This always sets the preview bit when acquiring a color scan.
The image quality may be reduced.
<p>
Some scanners (especally very old scanners) may need the options
<i>slow-speed</i> and/or <i>care-about-smearing</i>. The value
-1 uses the value defined by the driver, 0 disables the feature,
1 enables the feature. <b>Do not change these options unless
you really know what you do, you can destroy your scanner
by setting wrong values for this options!!!</b>
<p>
Some scsi drivers (or may be some scanners) do return an invalid
error code when a scsi command is executed. There are different
ways to handle such bad error codes. You can define the behaviour
of the umax-backend by setting the option <i>handle-bad-sense-error</i>
in umax.conf. You can set the following values:<br>
<dl>
<dd>0 : handle as device busy</dd>
<dd>1 : handle as ok</dd>
<dd>2 : handle as i/o error</dd>
<dd>3 : ignore bad error code, continue sense handler</dd>
</dl>
In most cases it does work when this condition is handled as a
device busy state so this is the default (0).
<p>
The option <i>execute-request-sense</i> disables(0) or enables(1)
the scsi command REQUEST_SENSE. The command can be called when the
scanner asks the driver to do calibration. The driver does
not know the required calibration parameters for all scanners.
Some scanners return the needed parameters when the driver
executes the REQUEST_SENSE command. But there is a problem
with the REQUEST_SENSE command on some systems with some
scsi controller drivers. This can cause a system hang.
So the safe way is to disable the REQUEST_SENSE command,
but the driver may not be able to do correct calibration
in this case.
<p>
For some scanners the calibration has to be done by the driver.
For this the driver has to know some values about the calibration
data. When the REQUEST_SENSE command is disabled or the scanner
does not return the correct data the driver needs to know the
correct parameters. For some scanners the parameters are stored
in the driver. When these values are not set correct the image
might look very strange. To be sure that you have a problem
with the calibration please enable debug output.
<p>
<i>calibration-full-ccd</i>:<br>
The calibration can be done for each element of the CCD (1)
or for each pixel of the requested image(0).
The value -1 enables the value defined in the driver.
<p>
<i>calibration-width-offset</i>:<br>
Some scanners use more pixels than calculated by image/CCD
parameters. A value larger or equal 0 defines the offset.
The value -1 enables the value defined in the driver.
<p>
<i>calibration-bytes-pixel</i>:<br>
Some scanners use 1 byte calibration data although the
scanner supports more than 8 bits/pixel. You can define
1 or 2 bytes per pixel calinration data.
The value -1 enables the value defined in the driver.
<p>
<i>invert-shading-data</i>:<br>
Some scanners need inverted shading data (1), others
need original shading data (0).
The value -1 enables the value defined in the driver.
</blockquote>
<h2>
SCSI ADAPTER TIPS:</h2>
<blockquote>
The UMAX-scanners do block the scsi-bus for a few seconds while
scanning. It is not necessary to connect the scanner to its own SCSI-adapter.
But if you need short response time for your SCSI-harddisk (e.g. if your
computer is a file-server), I suggest you use an own SCSI-adapter for your
UMAX-scanner.
</blockquote>
<blockquote>
The SCSI-adapters that are shipped with some Umax-scanners
are very simple ones. I suggest not to use them. If you really want to
try: the UDS-IS11 uses a DTC3181E chip (compatible to NCR5380). On linux
you may have the chance to get it work with the recent g_NCR5380-driver.
The card does not have IRQ/DMA. Try the following option for kernel boot
option:
<p> dtc_3181e=0x280,255
<p> or use the following command:
<p> insmod g_NCR5380 ncr_irq=255
ncr_addr=0x280 dtc_3181e=1
<p> - ncr_irq=255 : no IRQ, no DMA
<br> - ncr_irq=254 : autoprobe IRQ
<br> - ncr_addr : can be: 0x220 0x240 0x280 0x2A0 0x2C0 0x300 0x320 0x340
<br>
<p>
For
<a href="http://support.umax.co.uk/technotes/f096B.htm">information about the umax-scsi-cards</a>
take a look here.
</blockquote>
<blockquote>
See the relevant hardware FAQs and HOWTOs for your platform
for more information.
<br>See also: <a href="http://www.sane-project.org/man/sane-scsi.5.html">sane-scsi</a>
</blockquote>
<h2>FILES:</h2>
<blockquote>
<ul>
<li>The backend configuration file: /usr/local/etc/sane.d/umax.conf</li>
<li>The static library implementing this backend: /usr/local/lib/sane/libsane-umax.a</li>
<li>
The shared library implementing this backend : /usr/local/lib/sane/libsane-umax.so<br>
(present on systems that support dynamic loading)
</li>
</ul>
</blockquote>
<h2>ENVIRONMENT:</h2>
<blockquote>
<ul>
<li>SANE_DEBUG_UMAX</li>
</ul>
</blockquote>
<blockquote>
If the library was compiled with debug support enabled, this
environment variable controls the debug level for this backend. E.g., a
value of 128 requests all debug output to be printed. Smaller levels reduce
verbosity:
</blockquote>
<center>
<table BORDER>
<tr BGCOLOR="#40E0D0">
<td VALIGN=TOP COLSPAN="2"><b><font face="Arial">SANE_DEBUG_UMAX values</font></b></td>
</tr>
<tr>
<td WIDTH="10%">
<address>
<b>Number</b></address>
</td>
<td WIDTH="90%">
<address>
<b>Remark</b></address>
</td>
</tr>
<tr>
<td>
<address>
0</address>
</td>
<td>print important errors (printed each time)</td>
</tr>
<tr>
<td>
<address>
1</address>
</td>
<td>print errors</td>
</tr>
<tr>
<td>2</td>
<td>print sense</td>
</tr>
<tr>
<td>
<address>
3</address>
</td>
<td>print warnings</td>
</tr>
<tr>
<td>4</td>
<td>print scanner-inquiry</td>
</tr>
<tr>
<td>
<address>
5</address>
</td>
<td>print information</td>
</tr>
<tr>
<td>6</td>
<td>print less important information</td>
</tr>
<tr>
<td>7</td>
<td>print called procedures</td>
</tr>
<tr>
<td>8</td>
<td>print reader_process messages</td>
</tr>
<tr>
<td>
<address>
10</address>
</td>
<td>print called sane-init-routines</td>
</tr>
<tr>
<td>
<address>
11</address>
</td>
<td>print called sane-procedures</td>
</tr>
<tr>
<td>
<address>
12</address>
</td>
<td>print sane-infos</td>
</tr>
<tr>
<td>
<address>
13</address>
</td>
<td>print sane option control messages</td>
</tr>
</table></center>
<h4>Example:</h4>
<blockquote>
export SANE_DEBUG_UMAX=8
</blockquote>
<center>
<hr WIDTH="100%">
<br>
<font color="#009900">This page has been changed on 27th. feburary 2001</font>
<br>
<br>
</center>
</blockquote>
</td>
</tr>
</table>
<hr WIDTH="100%">
<blockquote>
Author: <a href="mailto:Oliver.Rauch@rauch-domain.de">Oliver Rauch</a>
</blockquote>
</body>
</html>
|